X-RAY DIFFRACTION CHARACTERIZATION OF CRYSTALLINITY AND PHASE COMPOSITION IN PLASMA-SPRAYED HYDROXYLAPATITE COATINGS

CHAPTER III
EXPERIMENTAL METHOD




X-ray diffraction data were acquired using copper Kα radiation, a graphite diffracted beam monochromator, and scintillation detector on a Bragg-Brentano focusing diffractometer. The x-ray source was a conventional sealed 2500 watt x-ray tube operated at 50kV and 40 mA. Incident and diffracted beam slits were 1.0 and 0.2 deg., respectively, chosen for high intensity. Data were collected by step counting at 0.02 deg. intervals for 2 seconds per data point.

3.1 Experimental Determination of Reference Intensity Ratios
RIR values were determined using commercially available pure samples of HA, β-TCP and CaO. Pure α-TCP could not be obtained, and only a mixture
containing a small fraction of β-TCP was available. The RIR value for α-TCP was obtained from the mixture of α- and β-TCP after quantification of the amount of β-TCP in the mixture. Alpha-quartz, rather than α-Al2O3, was selected initially as a suitable internal standard because peak interference was minimal and finely ground (5 μm) pure powder is readily available as a stable reference standard. Mixtures were prepared for measurement of the intensity ratios needed for the calculation of RIR values for each phase. At least five runs were made for each phase with the powder samples remixed and repacked in the sample holders between each run to reduce errors due to inhomogeneity of the powder mixtures. The mean and standard deviation for RIR values calculated for each phase are presented in Table 1.


 










The RIR values of a commercial sintered HA plasma spray starting powder and a laboratory precipitate preparation were found to differ by approximately 6%. Both samples were free of any contaminants detectable by x-ray diffraction and were obtained as “pure” HA powders. Minor variation or substitution in the apatite structure is assumed to alter the structure factor and the RIR values measured for the two samples. It is generally recommended that RIR values be determined experimentally for any phase using the data reduction methods and instruments to be employed for
measurement. [21]

3.2 Data Reduction Strategy
Known mixtures of HA, α-TCP, β-TCP, and CaO were prepared to develop a data collection and reduction strategy in which the angular regions of peak
integration and background were selected. The known mixtures were also used to test the validity of the RIR values and relative intensities. A typical diffraction pattern for a mixture containing 85% HA and additional contaminant phases, as indicated, is shown in Figure 2.




Plasma-sprayed HA coatings can vary widely in the amount of the glassy Ca/P phase producing the broad amorphous peak near 30 deg. The broad, diffuse peak from the glassy phase was approximated by fitting a Pearson VII function profile over a range sufficient to define the contribution of the broad tails. Much narrower Pearson VII functions were used to define the contribution of the neighboring HA peaks to the combined diffraction pattern. Peak profile fitting of the height, position, width and the power term independently for each peak was performed using the method of Gupta and Cullity [26] assuming a Kα doublet. The deconvolution of the overlapping peaks is shown in Figure 3.

After examination of the diffraction patterns of both the known mixtures and actual coating samples, regions of true background, beyond the tails of the broad glassy Ca/P peak, were selected which would be suitable for analysis of any likely concentrations of the phases encountered in plasma sprayed HA coatings. The following data reduction strategy was developed for acquiring the integrated intensities necessary to quantify mixtures of α-TCP, β-TCP, CaO, and HA. (Provision is made for interference effects from β-Ca2P2O7, but this phase has not been observed in plasma-sprayed coatings in this laboratory, and RIR values have yet to be determined.)
1.     The α-TCP content is obtained by integrating the α-TCP (331), (132), and (151) peaks from 23.8 to 24.8 degrees. This region is free from interference from any other phase likely to be found in HA samples. A quadratic background is assumed.
2.     The (008), (210), and (211) β-Ca2P2O7 peaks are integrated from 29.3 to 30.20 degrees to allow calculation of the intensity of the 100% peak. No pure sample of this phase has been available to allow the development of RIR values for quantification, but the integrated intensities may be corrected for interference. The integrated intensity will be in error due to interference from the β-TCP (300) and the α-TCP (113) peaks. This region is integrated assuming a linear background and a Pearson VII functional form of the peaks
3.     surrounding the region, including the diffuse peak 30.5 degrees to 31.5 degrees. The region in corrected for α-TCP (170) and (511) peak interference, and possible β-Ca2P2O7 (204) and (212) interference, if present. The β-TCP peak being used for quantification is the (0210) peak. This region is integrated in the same fashion as the β-Ca2P2O7 peaks above.
4.     The calcium oxide content is determined by integration from 37.0 degrees to 38.5 degrees with correction for the β-TCP (1211) and (315) peaks. This region contains the 100% (200) calcium oxide peak and is integrated assuming a quadratic form to the background.
5.     Finally, the entire region from 38.5 degrees to 59.0 degrees is integrated assuming a linear background and corrected for interference by α-TCP, β-TCP, calcium oxide, and possibly β-Ca2P2O7 to obtain the HA content. Many peaks are included in the integration range to reduce the effects of preferred orientation.





3.3 Relative Intensities
The relative intensities of the peaks or peak regions used in the analysis were obtained by direct measurement of samples of each phase. The integrated intensities were calculated from Pearson VII function peak profiles fitted by regression after background subtraction. Table 2 summarizes the combined fractional relative intensities for each phase summing all of the peaks found within the integration ranges shown. The HA values are the averages of results from pure samples obtained from five different sources.


 





















3.4 Plasma-sprayed coatings

HA plasma-sprayed coating samples reportedly obtained from a variety of commercial sources and representing a wide range of both dental and prosthetic applications, as well as experimental coatings. The set of coating samples was provided by a client for x-ray diffraction characterization and demonstration of the proposed method of quantitative phase analysis. The
origin of the individual samples was not provided to insure a blind evaluation. The results are reported here to show application of the method to a variety of commercial coatings and to demonstrate performance on a wide range of compositions. All the HA coating samples were analyzed nondestructively, mounting the coated substrates in a flat sample holder. A 1.0 deg. incident beam was used in a constant incident angle mode, allowing the width of the irradiated area to decrease with increasing diffraction angle. Using Cu Kα radiation and a graphite monochromator, as detailed above, the irradiated area ranged from nominally 10 x 12 mm at the lowest diffraction angle of 20 deg. to 10 x 5 mm at the highest angle of 60 deg. A quartz external standard was measured concurrently using the same experimental conditions.
The results for the plasma sprayed coating samples are presented in Table 4 and graphically in Figure 4. The crystallinity, taken as the sum of the crystalline fractions of the samples, varied from 38% to 84%. The most common secondary phase was β-TCP, which was observed in all of the coating samples, ranging from nominally 2% to 6%. Up to 2% α-TCP was attributed to four of the samples, but as noted above, the combination of the two TCP components is more reliable. No significant amount of CaO was observed in any of the coatings.


 















The error calculated for the determination of the major HA fraction in the coatings is higher (ranging from 2% to 6% for the individual samples) than that reported for the known mixtures (0.6% and 1.6%) in Table 3.



Diffraction from the glassy Ca/P phase, which exceeded 60% of the material in some of the coatings, must be subtracted form the contribution of the other phases introducing error in the determination of the net integrated intensities of all of the phases. The glassy phase does not occur in the fully crystalline known mixtures. The fraction of HA is calculated by subtracting the contributions from all of the other phases from the extended integration range, compounding the error in measurement.

A deficiency of the data reduction scheme currently employed is the sensitivity to the contribution of the secondary maximum of the amorphous glassy material to the long integration range (38.5 to 59.0 deg.) used to quantify the HA component. Integration of the full diffraction pattern obtained from the amorphous material produced by plasma spray deposition (Figure
1.g) reveals that the secondary maximum is on the order of 12% of the intensity of the primary, using the integration ranges proposed. For a typical HA plasma sprayed coating with a crystalline HA content on the order of 70%, the error due to the secondary maximum would be on the order of 4% of the HA content reported, less than the nominal random error for measurement of coatings. Because the intensity of the primary maximum is measured in the analysis, the method could be improved and applicability extended to coatings of lower crystallinity by correcting for the interference from the secondary amorphous maximum.